AlcoQuant 6020 plus
20 Doc no.: 059-07-1001779_A6020plus_GB-5.2.1 / 12.2021
How to insert the mouthpiece
The prepared original mouthpiece must be inserted into the mouthpiece port of the
AlcoQuant 6020 plus before measuring is performed. For this purpose the hygienic
packaging of the mouth piece must first be removed. Please use the hygienic packaging as
an hygienic protection (refer to pictures).
Possible faults Cancel measurement
If the blowing-in process is interrupted before sample collection is completed, a long dual
tone can be heard and an error message is displayed. The blowing-in process must be
repeated.
For every measurement please use a new mouth piece
The expiration aperture must be kept free during the blowing-
in process,
otherwise the measured value is distorted.
If it is suspected that the test subj
ect is under the influence of alcohol despite a
zero rea
ding, on account of his or her behavior or the smell of alcohol on his or
her breath, the device must be checked by trained service personnel.
Patients with reduced lung capacity may not achieve the r
equired expiration air of
1.2 l. In this case, a measure
ment is to be performed in the passive measurement
mode (without mouthpiece).
Start active mode measurement
After switch-on the device operates automatically in the
active mode and the Display shows “Start Measure“.