AlcoQuant 6020 plus
32 Doc no.: 059-07-1001779_A6020plus_GB-5.2.1 / 12.2021
Switching on
Error
Calibration
Confirm
with key
RED Device must be recalibrated.
No measurements are possible.
Measurement Error sound
Return to menu
Error
sound
Return to
menu
RED,
remains on
permanently
or RED
Errors prevent measurement
readiness.
No measurements are possible.
Contact service.
Measurement Error
Sampling
system
The sampling system cannot be
set into initial position.
Replace batteries; if problem
persists the device is faulty.
Contact service.
concen-tration
too high
Error
sound
Return to
menu
RED High concentration of alcohol
from previous measurement or
alcohol residue and moisture in
the blow-in tract.
Wait for a few minutes.
Blow through/dry the device with
a new mouthpiece.
Replace mouthpiece after every
measurement.
cally
measure
ment
readines
RED
Blow-in was too strong or too
weak. Blow into the device
slowly and constantly until the
acoustic beep stops.
Measurement Error
Sampling
system
No
measure
ment
value is
displayed
RED Sampling system not in nominal
position after measurement.
Hold the device still during the
sampling system and analysis.
Replace batteries; if problem
persists the device is faulty.
Contact service.
Measurement
Error
Temperature
No
measure
ment
value is
YELLOW
The device temperature was not
within the permitted range at the
time of analysis.