Sensepoint XCD RFD Technical Manual SPXCDHMRFENU Issue 1
28
(ZERO CALIBRATION)
1. If the ambient air is NOT considered reliable to use to set the ZERO, then remove
the weather protection and t the Gassing Cap accessory (see Section 4.3) onto the
sensor and apply a clean source of zero gas or compressed air.
2. To access the calibration menu, hold the end of the magnet over the switch located
at the top center of the detector display (
) for at least 3 seconds and then remove.
3. The display will indicate the rst conguration mode menu ‘SEt CAL’.
4. Put the magnet over the ‘
switch again and move to enter the Calibration menu.
5. The display will show the current gas reading, and the ‘ ’ icon ashes.
6. When the zero gas reading is stable use ‘
to conrm zero calibration.
7. If successful the display shows ‘ZEro PASS’ (if not successful, the display shows ‘ZEro
FAIL and returns to conguration mode).
8. If using zero-air, turn it off. Zeroing is complete and saved.
9. The display shows ‘SPAn’ with ‘YES’ ashing.
10.If span calibration is required use ‘
proceed to the next step. If span calibration is
not required, use ‘

to select ‘No’ and ‘
to return to conguration mode.