Sensepoint XCD RFD Technical Manual SPXCDHMRFENU Issue 1
41
bumP
tESt
PEAK READING
DISPLAYED WITH OUTPUT
INHIBITED ‘ ICON AND
PEAK STRING FLASHING.
APPLY BUMP TEST
GAS AND CHECK PEAK
READING ON DISPLAY.
CURRENT GAS
CONCENTRATION
DISPLAYED WITH
OUTPUT INHIBITED ‘
ICON FLASHING.
UNIT RETURNS TO
MENU MODE
Forc
Curr
‘Forc’ DISPLAYED WITH ‘
ICON. DEFAULT FORCE
CURRENT ‘4.00’ FLASHES.
USE

TO CHANGE TO
REQUIRED mA LEVEL.
UNIT TRANSMITS THE
CURRENT
IF YOU WANT TO EXIT
FROM THIS MENU, USE
ABORT FUNCTION
SEt
ALrm
‘AL1’, ICON DISPLAYED
AND CURRENT ALARM
LEVEL 1 CONCENTRATION
FLASHES. USE

TO
CHANGE TO REQUIRED
CONCENTRATION LEVEL.
‘AL1’ DISPLAYED WITH
‘NonE’, ‘rISE’ OR ‘FALL
FLASHING. USE

TO SELECT REQUIRED
ALARM ACTION.
‘AL2’, ICON
DISPLAYED
AND CURRENT
ALARM LEVEL 2
CONCENTRATION
FLASHES. USE

TO CHANGE
TO REQUIRED
CONCENTRATION
LEVEL.
‘AL2’ DISPLAYED WITH
‘NonE’, ‘rISE’ OR ‘FALL
FLASHING. USE

TO SELECT REQUIRED
ALARM ACTION AND USE
TO RETURN TO MENU
MODE.
Set
rLY
‘rL1’ DISPLAYED AND
‘AL1’, ‘AL2’, ‘Inht’ or ‘FLt’
FLASHES. USE

TO
CHANGE TO REQUIRED
RELAY TARGET.
‘rL1’ DISPLAYED
AND ‘dEEn’ or ‘Enr9’
FLASHES. USE

TO
CHANGE TO REQUIRED
RELAY ACTION.
‘Rl2’ DISPLAYED AND
‘AL1’, ‘AL2’, ‘Inht’ or
‘FLt’ FLASHES. USE

TO CHANGE TO
REQUIRED RELAY
TARGET.
‘rL2’ DISPLAYED AND
‘dEEn’ or ‘Encr9’ FLASHES.
USE

TO CHANGE
TO REQUIRED RELAY
ACTION.
‘rL3’ DISPLAYED AND
‘AL1’, ‘AL2’, ‘Inht’ or ‘FLt’
FLASHES. USE

TO
CHANGE TO REQUIRED
RELAY TARGET.
‘rL3’ DISPLAYED
AND ‘dEEn’ or ‘Enr9’
FLASHES. USE

TO
CHANGE TO REQUIRED
RELAY ACTION.
UNIT RETURNS TO
MENU MODE
rlY
OPr
‘rLY’, ‘ON’ DISPLAYED AND
CURRENT RELAY-ON TIME
FLASHES. USE

TO
CHANGE TO REQUIRED
RELAY-ON DELAY.
‘rLY’, ‘OFF’ DISPLAYED
AND CURRENT RELAY-
OFF TIME FLASHES.
USE

TO CHANGE
TO REQUIRED RELAY-
OFF DELAY.
‘Ltch’ DISPLAYED
AND ‘YES’ or ‘No’
FLASHES. USE

TO CHANGE TO
REQUIRED LATCH
OPTION.
UNIT RETURNS TO MENU
MODE
Set
Loc
‘Loc’ DISPLAYED WITH
FIRST 4 CHARACTERS OF
THE LOCATION STRING.
USE
TO CHANGE THE
1ST CHARACTER OF
CURRENT LOCATION
STRING. USE
TO MOVE
TO NEXT CHARACTER
AND SET REST OF
STRING. MAXIMUM 12
CHARACTERS CAN BE
SET.
‘Loc’ DISPLAYED NEW
LOCATION STRING. THE
STRING MOVES RIGHT-
TO-LEFT TO SHOW
WHOLE CHARACTERS
TWICE. THEN UNIT
AUTOMATICALLY
RETURNS TO MENU
MODE.
tEm
PUnIt
‘tEmP’ DISPLAYED WITH
‘°C’ OR ‘°F’ FLASHING.
USE

TO CHANGE TO
REQUIRED TEMPERATURE
UNIT.
UNIT RETURNS TO
MENU MODE
tES
tALrm
‘Forc’ DISPLAYED and ’AL1’
FLASHING TO SELECT
TEST-REQURED ALARM.
USE ‘
TO FORCE
ALARM1.
GAS NAME, CURRENT
GAS CONCENTRATION
DISPLAYED AND
ICON AND AL ICON
FLASH.
UNIT RETURNS TO
MENU MODE
9uIt
QUITS MENU MODE AND
RETURNS TO MONITORING
MODE

OK

OK

OK

OK

1
Refer to section 10.1 for detailed zero and span calibration procedures. Re-calibrate the detector if left un-powered for
periods in excess of 24 hours.
2
The calibration due warning counter is automatically reset after a successful calibration.
Table 12. Conguration Mode Operation