PHASe:UNLock:ERRor:STATe {OFF|ON}
Disable or enable the function generator from generating an error
if phase-lock is ever lost. If phase-lock is lost and the error is enabled,
580, “Phase-locked loop is unlocked is generated. The default is
OFF.
[ Stored in non-volatile memory ]
PHASe:UNLock:ERRor:STATe?
Query the unlock error state. Returns “0” (
OFF) or “1” (ON).
OUTPut:TRIGger:IMMediate
Output an immediate
TTL “high” pulse from the rear-panel Ext Trig
terminal regardless of the present setting of the OUTP:TRIG:STAT
command. You can use this command to issue an immediate external
trigger for synchronizing phase-lock signals using the rear-panel
Ext Trig terminal.
OUTPut:TRIGger:STATe {OFF|ON}
Disable or enable the function generator from sourcing an external
trigger from its rear-panel Ext Trig terminal. The default is
OFF.
[ Stored in volatile memory ]
OUTPut:TRIGger:STATe?
Query the external trigger state. Returns “0” (
OFF) or “1” (ON).
*OPT?
Query the presence of the Phase-Lock option. Returns “1:
PLL” if the
option is present or “0” if no option is present.