Autozero
(continued)
The following table shows the relationship between integration time and
autozero settings from the remote interface and the corresponding
front-panel settings.
From HP 34401A Specifications:
1
Remote Configuration
NPLC: 100
Autozero: On
Digits Displayed: 6
1
2
NPLC: 100
Autozero: Off
Digits Displayed: 6
1
2
NPLC: 10
Autozero: On
Digits Displayed: 6
1
2
NPLC: 10
Autozero: Off
Digits Displayed: 6
1
2
NPLC: 1
Autozero: On
Digits Displayed: 5
1
2
NPLC: 1
Autozero: Off
Digits Displayed: 5
1
2
NPLC: 0.2
Autozero: On
Digits Displayed: 5
1
2
NPLC: 0.2
Autozero: Off
Digits Displayed: 5
1
2
NPLC: 0.02
Autozero: On
Digits Displayed: 4
1
2
NPLC: 0.02
Autozero: Off
Digits Displayed: 4
1
2
Front-Panel Equivalent
Slow 6 digits
N/A
Fast 6 digits
Slow 5 digits
N/A
Slow 4 digits
N/A
N/A
Fast 5 digits
N/A
Fast 4 digits
Digits Displayed
N/A
6
1
2
N/A
6
1
2
N/A
5
1
2
N/A
5
1
2
N/A
4
1
2
Readings/Sec
N/A
0.6
N/A
6
N/A
60
N/A
300
N/A
1000
1
See the HP 34401A specifications listed on page 217.
Chapter 3 Features and Functions
Measurement Configuration
60