Automatic Trigger Delays
If you do not specify a trigger delay, the multimeter selects an
automatic delay for you. The delay is determined by function, range,
integration time, and ac filter setting.
DC Voltage and DC Current (for all ranges):
Resistance (2-wire and 4-wire):
AC Voltage and AC Current (for all ranges):
Frequency and Period:
Integration Time
NPLC 1
NPLC < 1
Trigger Delay
1.5 ms
1.0 ms
Range
100
1 k
10 k
100 k
1 M
10 M
100 M
Trigger Delay
(For NPLC
1)
1.5 ms
1.5 ms
1.5 ms
1.5 ms
15 ms
100 ms
100 ms
AC Filter
Slow
Medium
Fast
Trigger Delay
7.0 sec
1.0 sec
600 ms
Remote or single/external trigger
Range
100
1 k
10 k
100 k
1 M
10 M
100 M
Trigger Delay
(For NPLC < 1)
1.0 ms
1.0 ms
1.0 ms
1.0 ms
10 ms
100 ms
100 ms
AC Filter
Slow
Medium
Fast
Trigger Delay
1.5 sec
200 ms
100 ms
Front panel with auto trigger ON
Trigger Delay
1.0 sec
Remote or single/external trigger
Trigger Delay
0 sec
Front panel with auto trigger ON