EasyManuals Logo

HP 34970A User Manual

HP 34970A
429 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #130 background imageLoading...
Page #130 background image
Remote Interface Operation (Digital Input Channel): To assign the
alarm number to report any alarm conditions on the specified digital
input channels, use the following command.
OUTPut:ALARm[1|2|3|4]:SOURce (
@<ch_list>)
To configure alarms on the specified digital input channel, use
the following commands (also see the example on the following page).
CALCulate
:COMPare:TYPE {EQUal|NEQual}[,(
@<ch_list>)]
:COMPare:DATA <
data>[,(@<ch_list>)]
:COMPare:MASK <
mask>[,(@<ch_list>)]
Select EQUal to generate an alarm when the data read from the port
is equal to CALC:COMP:DATA after being masked by
CALC:COMP:MASK.
Select NEQual (not equal) to generate an alarm when the data read
from the port is not equal to CALC:COMP:DATA after being masked by
CALC:COMP:MASK.
Use CALC:COMP:MASK to designate the “don’t care” bits. Bits that you
set to “0” in the mask are ignored.
To enable the specified alarm mode, send the following command.
CALCulate:COMPare:STATe ON [,(
@<ch_list>)]
4
Chapter 4 Features and Functions
Alarm Limits
131