Frequency Configuration Commands
See also “Frequency Measurement Configuration” in chapter 4 starting on
page 118.
[SENSe:]
FREQuency:RANGe:LOWer {3|20|200|MIN|MAX}[,(
@<ch_list>)]
FREQuency:RANGe:LOWer? [{(
@<ch_list>)|MIN|MAX}]
Specify the lowest frequency expected in the input signal for frequency
measurements on the specified channels. The instrument selects a slow,
medium (default), or fast measurement timeout based on the frequency
you specify.
MIN = 3 Hz. MAX = 200 Hz. The :LOW? query returns the
timeout setting on the specified channels. Returns “3”, “20”, or “200”.
5
Chapter 5 Remote Interface Reference
Frequency Configuration Commands
225