EasyManuals Logo

HP 34970A User Manual

HP 34970A
429 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #307 background imageLoading...
Page #307 background image
Instrument Errors
111 Channel list: slot number out of range
The specified slot number is invalid. The channel number has the
form (
@scc), where s is the slot number (100, 200, or 300) and cc is the
channel number. Example: CONF:VOLT:DC (
@404)
112 Channel list: channel number out of range
The specified channel number is invalid for the module in the selected slot.
The channel number has the form (
@scc), where s is the slot number
(100, 200, or 300) and cc is the channel number.
Example: ROUT:CLOSE (
@134)
113 Channel list: empty scan list
Before you can initiate a scan, you must set up a scan list which includes
all configured multiplexer or digital channels in the instrument. Use the
MEASure?, CONFigure, or ROUTe:SCAN commands to set up your scan list.
201 Memory lost: stored state
This error is reported at power-on to indicate that a stored state has
become unusable. This error is most likely caused by a dead battery
(memory is battery-backed). Refer to the HP 34970A Service Guide to
replace the internal battery.
202 Memory lost: power-on state
This error is reported at power-on to indicate that the power-down state
of the instrument (normally recalled when power is turned on) has become
unusable. This error is most likely caused by a dead battery (memory is
battery-backed). Refer to the HP 34970A Service Guide to replace the
internal battery.
203 Memory lost: stored readings
This error is reported at power-on to indicate that readings stored in
memory from a previous scan have been lost. This error is most likely
caused by a dead battery (memory is battery-backed). Refer to the
HP 34970A Service Guide to replace the internal battery.
204 Memory lost: time and date
This error is reported at power-on to indicate that the time and date
settings have been lost (they are reset to
JAN 1, 1996 00:00:00).
This error is most likely caused by a dead battery (memory is
battery-backed). Refer to the HP 34970A Service Guide to replace the
internal battery.
6
Chapter 6 Error Messages
Instrument Errors
309

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the HP 34970A and is the answer not in the manual?

HP 34970A Specifications

General IconGeneral
BrandHP
Model34970A
CategorySwitch
LanguageEnglish

Related product manuals