[SENSe:]
VOLTage:DC:RANGe {<
range>|MIN|MAX}[,(@<ch_list>)]
VOLTage:AC:RANGe {<
range>|MIN|MAX}[,(@<ch_list>)]
RESistance:RANGe {<
range>|MIN|MAX}[,(@<ch_list>)]
FRESistance:RANGe {<
range>|MIN|MAX}[,(@<ch_list>)]
CURRent:DC:RANGe {<
range>|MIN|MAX}[,(@<ch_list>)]
CURRent:AC:RANGe {<
range>|MIN|MAX}[,(@<ch_list>)]
FREQuency:VOLTage:RANGe {<
range>|MIN|MAX}[,(@<ch_list>)]
PERiod:VOLTage:RANGe {<
range>|MIN|MAX}[,(@<ch_list>)]
Select the measurement range for the function selected on the specified
channels.
MIN selects the lowest range. MAX selects the highest range.
[SENSe:]
VOLTage:DC:RANGe? [{(
@<ch_list>)|MIN|MAX}]
VOLTage:AC:RANGe? [{(
@<ch_list>)|MIN|MAX}]
RESistance:RANGe? [{(
@<ch_list>)|MIN|MAX}]
FRESistance:RANGe? [{(
@<ch_list>)|MIN|MAX}]
CURRent:DC:RANGe? [{(
@<ch_list>)|MIN|MAX}]
CURRent:AC:RANGe? [{(
@<ch_list>)|MIN|MAX}]
FREQuency:VOLTage:RANGe? [{(
@<ch_ list>)|MIN|MAX}]
PERiod:VOLTage:RANGe? [{(
@<ch_list>)|MIN|MAX}]
Query the measurement range on the specified channels. Returns a
number in the form ā€œ
+1.00000000E+01ā€.
[SENSe:]
VOLTage:DC:RANGe:AUTO {OFF|ON}[,(
@<ch_list>)]
VOLTage:AC:RANGe:AUTO {OFF|ON}[,(
@<ch_list>)]
RESistance:RANGe:AUTO {OFF|ON}[,(
@<ch_list>)]
FRESistance:RANGe:AUTO {OFF|ON}[,(
@<ch_list>)]
CURRent:DC:RANGe:AUTO {OFF|ON}[,(
@<ch_list>)]
CURRent:AC:RANGe:AUTO {OFF|ON}[,(
@<ch_list>)]
FREQuency:VOLTage:RANGe:AUTO {OFF|ON}[,(
@<ch_list>)]
PERiod:VOLTage:RANGe:AUTO {OFF|ON}[,(
@<ch_list>)]
Disable or enable autoranging on the specified channels. Autoranging
thresholds: Down ranges at <10% of range; Up range at >120% of range.