[SENSe:]
TOTalize:SLOPe {NEGative|POSitive}[,(
@<ch_list>)]
TOTalize:SLOPe? [(
@<ch_list>)]
Configure the totalizer to count on the rising edge (default; positive) or
falling edge (negative) of the input signal. The totalizer channel is
numbered s03”, where s represents the slot number. The :SLOP? query
returns the edge selection. Returns “
NEG” or “POS”.
[SENSe:]TOTalize:CLEar:IMMediate [(
@<ch_list>)]
Immediately clear the count on the specified totalizer channels.
The totalizer channel is numbered s03”, where s is the slot number.
[SENSe:]TOTalize:DATA? [(
@<ch_list>)]
Read the count on the specified totalizer channels. If you have selected
the RRESet parameter in the TOT:TYPE command, then the TOT:DATA?
command will reset the count to “0” after it is read. The count is reset
regardless of whether the specified channels are in the scan list or even
whether a scan is in progress. The maximum count is 67,108,863 (2
26
-1).
The count rolls over to “0” after reaching the maximum allowed value.
The output from this command is affected by the FORMat:READing
commands (see “Reading Format Commands” on page 231).