EasyManuals Logo

HP 34970A User Manual

HP 34970A
429 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #265 background imageLoading...
Page #265 background image
INSTrument
:DMM {OFF|ON}
:DMM?
Disable or enable the internal
DMM. When you change the state of the
internal
DMM, the instrument issues a Factory Reset (*RST command).
The :DMM? query returns the state of the internal DMM. Returns “0”
(disabled) or “1” (enabled).
INSTrument:DMM:INSTalled?
Query the instrument to determine if the internal
DMM is installed.
Returns “0” (not installed) or “1” (installed).
*RST
Reset the instrument to the Factory configuration. See “Factory Reset
State” on page 160 in chapter 4 for a complete listing of the instrument’s
Factory Reset state.
This command is equivalent to selecting
FACTORY RESET from the
front-panel Sto/Rcl Menu.
SYSTem:PRESet
Preset the instrument to a known configuration. See “Instrument Preset
State” on page 161 in chapter 4 for a complete listing of the instrument’s
preset state.
This command is equivalent to selecting
PRESET from the front-panel
Sto/Rcl Menu.
SYSTem:CPON {100|200|300|ALL}
Reset the module in the specified slot to its power-on state (CPON means
“card power on”). To reset all three slots, specify ALL.
This is equivalent to pressing from the front panel.
5
Chapter 5 Remote Interface Reference
System-Related Commands
267

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the HP 34970A and is the answer not in the manual?

HP 34970A Specifications

General IconGeneral
BrandHP
Model34970A
CategorySwitch
LanguageEnglish

Related product manuals