INSTrument
:DMM {OFF|ON}
:DMM?
Disable or enable the internal
DMM. When you change the state of the
internal
DMM, the instrument issues a Factory Reset (*RST command).
The :DMM? query returns the state of the internal DMM. Returns “0”
(disabled) or “1” (enabled).
INSTrument:DMM:INSTalled?
Query the instrument to determine if the internal
DMM is installed.
Returns “0” (not installed) or “1” (installed).
*RST
Reset the instrument to the Factory configuration. See “Factory Reset
State” on page 160 in chapter 4 for a complete listing of the instrument’s
Factory Reset state.
This command is equivalent to selecting
FACTORY RESET from the
front-panel Sto/Rcl Menu.
SYSTem:PRESet
Preset the instrument to a known configuration. See “Instrument Preset
State” on page 161 in chapter 4 for a complete listing of the instrument’s
preset state.
This command is equivalent to selecting
PRESET from the front-panel
Sto/Rcl Menu.
SYSTem:CPON {100|200|300|ALL}
Reset the module in the specified slot to its power-on state (CPON means
“card power on”). To reset all three slots, specify ALL.
This is equivalent to pressing from the front panel.