EasyManua.ls Logo

HP 34970A - Page 286

HP 34970A
429 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Standard Event Register Commands
See the table on page 282 for the register bit definitions.
*ESR?
Query the event register in this register group. This is a read-only
register. Once a bit is set, it remains set until cleared by a *CLS
(clear status) command. A query of this register returns a decimal
value which corresponds to the binary-weighted sum of all bits set in
the register.
*ESE <
enable_value>
*ESE?
Enable bits in the enable register in this register group. The selected
bits are then reported to the Status Byte. A *CLS (clear status) will not
clear the enable register but it does clear all bits in the event register.
A STATus:PRESet does not clear the bits in the Standard Event enable
register. To enable bits in the enable register, you must write a decimal
value which corresponds to the binary-weighted sum of the bits you wish to
enable in the register.
The *ESE? query returns a decimal value which corresponds to the
binary-weighted sum of all bits enabled by the *ESE command.
Chapter 5 Remote Interface Reference
Status System Commands
288

Table of Contents

Related product manuals