3-4
Performance Tests
Recall Default Settings
The Performance Tests require the HP 37718A/19A to be set to a pre-
defined (default) state at the beginning of each test.
1 Using , display softkeys, and set up the
display as shown opposite
.
2 Press to recall the instrument default settings.The
instrument display will blank for a few seconds while the settings are
recalled and the status display will indicate stored settings number 0
recalled.
OTHER
OTHER
STORED
SETTINGS
RECALL