Step
1. Initial
Observations
Initiate
the Analyzer
Self-Test
1.
Disconnect
all
devices
and p
eripherals from
the analyzer
(including all
test set
in
terconnects).
2.
Switc
h
on
the
analyzer
and
press
4
PRESET
5
.
3.
W
atc
h
for
the
indications sho
wn in
Figure 4-1
to determine
if
the
analyzer
is
op
erating
correctly
.
Figure
ORDER4
here.
Figure
4-1.
Preset
Sequence
If
the
Self-T
est
F
ailed
Chec
k the
AC
line
p
o
w
er
to
the
analyzer.
Chec
k
the
fuse
(rating
listed
on rear
panel, spare
inside holder).
Chec
k
the
line v
oltage setting
(use small
screwdriver
to c
hange).
If the problem persists, refer to \Step 4. F
aulty Group Isolation."
DRAFT
3/21/106 15:12
Start Troubleshooting Here 4-3