V
erify the
A15 Preregulator
Verify
that the
A15 preregulator
is supplying
the correct
voltages
to the
A8
p
ost
regulator.
Use
av
oltmeter with
a
small
prob
e
to
measure
the
output
v
oltages
of
A15W1's plug.
Refer to
T
able
5-2
and
Figure
5-6
.
If
the
voltages
are not
within tolerance,
replace A15.
If
the
v
oltages
are
within
tolerance,
A15
is
v
eried.
Con
tin
ue
to
\Chec
k
for
aF
aulty
Assem
bly
."
T
able
5-2.
Output
V
oltages
Pin A15W1P1
(Disconnected)
Voltages
A8J2
(Connected)
V
oltages
A15
Preregulator
Mnemonic
1,2 +125
to
+100
+68
to
+72
+70
V
3,4 +22.4 to
+33.6
+17.0 to
+18.4
+18 V
5,6
0
22.4
to
0
33.6
0
17.0
to
0
18.4
0
18
V
7 N/C N/C N/C
8 +9.4
to +14
+7.4
to +8.0
+8
V
9,10
0
9.4
to
0
14
0
6.7
to
0
7.3
0
8
V
11 +32
to
+48
+24.6
to
+26.6
+25
V
12 N/C N/C N/C
NOTE:
The
+5VD
supply
m
ust
b
e
loaded
b
yone
or
more
assem
blies
at
all
times,
or
the
other
v
oltages
will
not
b
e
correct.
It
connects
to
motherb
oard
connector
A17J3
Pin
4.
Figure
PLUG5
here.
Figure 5-6. A15W1 Plug Detail
5-8 Power Supply Troubleshooting DRAFT
3/21/106 15:12
Artisan Scientific - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisan-scientific.com

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the HP 8753D Option 011 and is the answer not in the manual?

HP 8753D Option 011 Specifications

General IconGeneral
BrandHP
Model8753D Option 011
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals