EasyManua.ls Logo

HP 8753D Option 011

HP 8753D Option 011
398 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
HP 8753D
Performance
Test
Record (10
of 18)
For
30
kHz-6
GHz
Analyzers
Hewlett-Pac
kard
Company
Mo
del HP
8753D Option
011 and
Option
006
Rep
ort Num
b er
Serial
Num
b
er
Date
7
7
8.
Phase
F
requency
Resp
onse
F
requency Range
Ratio Sp
ecication
Measured
Value
Measuremen
t
Uncertain
t
y
300 kHz
-3
GHz
A/R
6
3
6
0.67
300
kHz
-
3
GHz
B/R
6
3
6
0.67
300
kHz
-
3
GHz
A/B
6
3
6
0.67
3GHz
-6
GHz
A/B
6
10
6
0.67
3
GHz
-
6
GHz
A/R
6
10
6
0.67
3
GHz
-
6
GHz
B/R
6
10
6
0.67
2b-12 Performance Test Record DRAFT
3/21/106 15:21
Artisan Scientific - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisan-scientific.com

Table of Contents

Related product manuals