HP 8753D
Performance
Test
Record (14
of 18)
For
30
kHz|6
GHz
Analyzers
Hewlett-Pac
kard
Company
Mo
del HP
8753D Option
011 and
Option 006
Rep
ort Num
b er
Serial
Num
b
er
Date
7
7
12.
Receiv
er
Magnitude
Dynamic
Accuracy
G F
j
G
0
F
j
T
est
P
ort
Input
P
o
wer
8496A
A
ttn.
(dB)
T
est
P
ort
Measuremen
t
(dBm)
Exp
ected
Measuremen
t
(dBm)
Dynamic
Accuracy
(Calculated)
Sp
ec.
(dB)
Meas.
Uncer.
(dB)
Channel
B
0
10
0
0.075
6
0.008
0
20
(Ref
)
10 Reference Reference Reference Reference Reference
0
30
20
0.050
6
0.008
0
40
30
0.050
6
0.008
0
50
40
0.050
6
0.008
0
60
50
0.058
6
0.017
0
70
60
0.089
6
0.017
0
80
70
0.240
6
0.017
0
90
80
0.680
6
0.017
0
100
90
1.950
6
0.027
Channel
A
0
10
0
0.075
6
0.008
0
20
(Ref
)
10 Reference Reference Reference Reference Reference
0
30
20
0.050
6
0.008
0
40
30
0.050
6
0.008
0
50
40
0.050
6
0.008
0
60 50
0.058
6
0.017
0
70
60
0.089
6
0.017
0
80
70
0.240
6
0.017
0
90
80
0.680
6
0.017
0
100
90
1.950
6
0.027
2b-16 Performance Test Record DRAFT