50
AUTO
-LENSMETER
4.5. UV MODULE TEST
Check UV-LED states
UV-LED values in UV mode should change(100% -> 0%).
You can get UV values at most 100%.
(When you get in the UV mode, then popup message shows to adapt auto 100%
function)
Check1. Check if light radiated from OPTIC
LED comes out from UV measurement part.
NO
Change UV-LED
PCB ASS’Y
Check1. Check if UV SENSER in UV measurement part is hidden by
obstacles such as FILTER COVER.
NO
Check1. Turn on the power switch and enter UV MODE.
Check2. Turn variable resistance in UV-REC PCB with cross screw driver.
Check3. Press CAL BUTTON among MODE BUTTON and you can get 100%.
Change UV MODULE
NO
Change MAIN BOARD PCB ASS’Y