Auto Lensmeter HLM-7000P
53
5. How to upgrade the firmware software
5.1. Introduction
Upgrade software for HLM-7000P proceeds by serial communication. DNW which is PC
program to transmit firmware software from PC to Lensmeter is necessary.
An administrator of the Lensmeter hardware can receive the Firmware by e-mail or else
and load it to his/her lensmeter by this downloading program when the new upgrade
version of firmware comes out. If there is PC with Windows operating system and serial
cable connected to the Lensmeter, with simple process about 3 minutes, he/she can
upgrade the performance of his/her Lensmeter without sending and receiving it to the
agency.
This downloading program is DNW application.
Downloading process is checked by DNW display.
Check the following item and process to update software:
Check firmware version of the Lensmeter
Connect serial cable between PC and Lensmeter
Check COM port of the Lensmeter
Connect COM port using DNW
Power on Lensmeter with pressing 5 key
Transmit firmware binary file to the Lensmeter using DNW