Auto Lensmeter HLM-7000P
15
2.6. PD Setup
With the PD Sensor, we can calibrate the PD value of our machine.
The procedures are as followings:
A. Press the ‘PD’ ( ) button.
B. Push in the PD sensor is the leftmost. And press the ‘L Limbutton.
C. Push in the PD sensor is the rightmost. And press the ‘R Lim button and check if
‘PD POSITION[0] ~ PD POSITION[8]’ values are changed.
D. Finally the ‘SAVEbutton to save.
The information displayed in ‘setting the position of PDscreen is as follows: