17. Wearing gloves is recommended when performing all treatments, system cleaning and preventive maintenance activities.
18. Device cannot be serviced or maintained while in use.
19. Unauthorized user modications will void warranty. Edge Systems LLC will not be responsible for any injuries sustained
due to unauthorized System modications or use of parts not specied by Edge Systems LLC. Systems returned to Edge Systems LLC
with unsafe modications will be returned to their original operating condition at the customer’s expense.
20.
DO NOT suction liquids through the dry port located on the side of the unit. This will damage the System and void the
warranty.
21. Use of this equipment adjacent to or stacked with other equipment should be avoided because it could result in
improper operation. If such use is necessary, this equipment and the other equipment should be observed to verify that they are
operating normally.
22.
If this System does cause interference, which can be determined by turning the System off and on, the interference can
be possibly corrected by one or more of the following measures:
•
Relocate the System with respect to the receiver.
• Move the System away from the receiver.
• Plug the System into a different electrical outlet so that the System and receiver are on different branch circuits.
23. Use of accessories, transducers and cables other than those specied or provided by Edge Systems LLC could result in increased
electromagnetic emissions or decreased electromagnetic immunity of this equipment and result in improper operation.
Note: The emissions characteristics of this equipment make it suitable for use in industrial areas and hospitals (CISPR 11 Class A).
If it is used in a residential environment (for which CISPR 11 Class B is normally required), this equipment might not offer adequate
protection to radio-frequency communication services. The user might need to take mitigation measures, such as relocating or
reorienting the equipment.
Emissions Standard Test Method
EN 606061-1-2: 2015
IEC 60601-1-2:2014
CISPR 11:2015 +A1:2016
EN 61000-3-2:2014
EN 61000-3-3:2013
Immunity Standard Test Method
EN 60601-1-2:2015
IEC 60601-1-2:2014
IEC 61000-4-2:2008
IEC 61000-4-3:2010
IEC 61000-4-4:2012
IEC 61000-4-5:2014
IEC 61000-4-6:2013
IEC 61000-4-8:2009
IEC 61000-4-11:2004
ISO 7637-2
24. Portable RF communications equipment should be used no closer than 15 cm/6 in to any part of the HydraFacial
System, including cables specied by Edge Systems LLC. Otherwise, degradation of the performance of this equipment could
result.
25.
To prevent electric shock, do not remove the instrument cover. There are no user serviceable parts inside. Routine
maintenance or cleaning of internal parts is not necessary. Avoid the use of cleaning agents or chemicals on the instrument. Some
chemicals may damage plastic parts or lettering. Any external cleaning should be done with a clean, dry or slightly damp cloth.
Unpacking
Inspect the Shipping Carton
Your System was shipped in a custom foam insulated box that has been tested to UPS standards. If the shipping carton is damaged,
inspect the contents for visible damage such as dents, scratches or any other obvious signs of damage. If the System is damaged,
notify Edge Systems LLC or your local distributor and they will assist you in the repair or replacement of your System.
Do not return your unit without rst contacting your distributor and receiving an RMA (Return Material Authorization) number.
Unpack the unit, accessories and all documents. Retain the original packaging in case you need to ship the System for any reason.
Verify that all appropriate items have been received.
The unit is large and heavy and should not be lifted alone.
Package Contents
• Console Unit
• Plastic Cartridge Handpiece Assembly
• Tool Holder
• Puncture Tool
• Waste Canister Assembly
• Cleaning Caps, 5-Pack
• HydraFacial Power Cord (Region Specic)