This Manual Copyright © 2012 by Hygiena LLC
All rights reserved. No part of this manual may be reproduced,
stored in a retrieval system, or transmitted, by any means or in any
form, without the prior permission of Hygiena LLC.
The information contained in this manual is subject to change
without notice. Hygiena LLC assumes no responsibility for any errors
that may appear in this document.
This manual will describe the specifications and general operations
of the SystemSURE Plus v2 and EnSURE v2 luminometers.
For older luminometers refer to the operators manual that shipped
with the luminometer for specific operations.
Document: SystemSURE Operators Manual V3.0. 2013
Copyrights and Trademarks
Worldwide Patents Pending.
Hygiena, EnSURE, SystemSURE Plus, SystemSURE II,
Ultrasnap, SureTrend and dataSURE II are trademarks of
Hygiena LLC.
Operating Precautions and Limitations of Use
The following symbol is used in this manual:
Description: CAUTION / WARNING
IMPORTANT
Hygiena’s products are designed and constructed to be safe
and without risk to health when properly used in accordance
with the supplied documentation and when the operating
precautions outlined in this manual are fully observed.
It is essential that the user of this manual is aware of the
potential hazards associated with the unit and its accessories
All operators should be familiar with the safety precautions
and warnings given in this manual before attempting to
operate the unit.
If the unit is used in a manner which is not specified by the
manufacturer, the protection provided by the equipment may
be impaired.
Please read this manual in its entirety and the “Operating
Precautions and Limitations of Use” at the end of the manual
before using the Luminometer.

Questions and Answers

  • M
    moranstevenAug 22, 2025
    What to do if Hygiena Measuring Instruments measurement result is higher than expected?
    • T
      timothymcclainAug 22, 2025
      If the measurement result from your Hygiena Measuring Instruments is higher than expected, consider the following: Ensure correct and consistent use of the testing device. Verify that the lid is fully closed. Be aware that an unstable thermal environment may affect readings; try turning the luminometer off and back on. Also, check the luminometer casework or lid for damage or faults.
  • J
    Jason KaufmanSep 4, 2025
    What does E5 mean on Hygiena SystemSURE Plus v2?
    • A
      Amanda GreenSep 4, 2025
      Error E5 on your Hygiena Measuring Instruments indicates that program thresholds, test plan data, or the User ID are not defined.
  • S
    Steven McbrideSep 2, 2025
    What does E2 mean on Hygiena SystemSURE Plus v2?
    • W
      Wanda MontoyaSep 2, 2025
      Error E2 on your Hygiena Measuring Instruments indicates that the luminometer is being used or was stored outside of its specified operating temperature range, or that the luminometer is damaged or faulty.
  • J
    Julie PattonSep 1, 2025
    What does E3 mean on Hygiena SystemSURE Plus v2 Measuring Instruments?
    • B
      Bryan WilliamsSep 1, 2025
      Error E3 on your Hygiena Measuring Instruments indicates that the luminometer environment is unstable or being used in an area of high electromagnetic noise. It may also mean the luminometer was tilted during measurement, the lid is damaged allowing light in, the protective pocket is dirty or scratched, or the luminometer itself is damaged or faulty.
  • M
    Margaret HolderAug 30, 2025
    Why does Hygiena SystemSURE Plus v2 clock reverts to 12:00 01/01/2000?
    • K
      Kimberly BurgessAug 31, 2025
      If the clock on your Hygiena Measuring Instruments reverts to 12:00 01/01/2000, it could be due to the batteries having been replaced, low batteries, or loose batteries within the compartment. Additionally, the luminometer may have been dropped or subjected to shock or vibration, or the luminometer itself may be damaged or faulty.
  • K
    Kayla TylerAug 29, 2025
    What does E6 error mean on Hygiena SystemSURE Plus v2 Measuring Instruments?
    • M
      Melanie RiceAug 29, 2025
      If your Hygiena Measuring Instruments shows error E6, power the luminometer off and then back on. This error can also indicate failed calibration self-checks, an unstable operating environment, a dirty or scratched protective pocket, a lid that is not fully closed, a damaged lid seal, a damaged protective pocket, or a damaged or faulty luminometer.
  • H
    houstonbrittanyAug 27, 2025
    What to do if Hygiena Measuring Instruments luminometer will not turn off when the button is pressed?
    • A
      Adam HodgesAug 28, 2025
      If your Hygiena Measuring Instruments luminometer will not turn off, it might be busy performing a reading or calibration self-check operation. If this is not the case, try removing the batteries for 30 seconds and then reinserting them. It's also possible that the luminometer or keypad is damaged or faulty.
  • K
    Kimberly WardAug 27, 2025
    What to do if Hygiena Measuring Instruments measurement result is always zero RLU or is much lower expected?
    • J
      Joseph WeissAug 27, 2025
      If the measurement result from your Hygiena Measuring Instruments is always zero RLU or much lower than expected, consider the following: ensure correct use of the testing device, make sure the luminometer is upright when taking a measurement, check if the testing device is out-of-date, turn the luminometer off and back on if used in an unstable thermal environment, ensure the protective pocket or testing device is fully inserted, check the protective pocket (it might be dirty, severely scratched, or damaged), or the luminometer may be damaged or faulty.
  • I
    Ivan BennettAug 25, 2025
    Why Hygiena SystemSURE Plus v2 Measuring Instruments luminometer turns off unexpectedly?
    • J
      Jessica RichardsonAug 25, 2025
      Your Hygiena Measuring Instruments luminometer may be turning off unexpectedly because the batteries are low or loose. Other potential causes include if the luminometer was dropped or subjected to shock/vibration, if the luminometer has not been used for 10 minutes (automatically entering standby mode), or if the luminometer is damaged or faulty.
  • J
    Jeremy WarrenAug 23, 2025
    What to do if Hygiena SystemSURE Plus v2 luminometer will not turn on when the button is pressed?
    • M
      Meredith GonzalezAug 23, 2025
      If your Hygiena Measuring Instruments luminometer won't turn on, check if the batteries are low, of the wrong type, or incorrectly inserted. It's also possible that the luminometer or keypad is damaged or faulty.

Summary

Operating Precautions and Limitations of Use

1. Introduction

1.2 How Does the Luminometer Work?

Explains the chemical reaction and light detection process in the luminometer.

2. Luminometer Description

2.2 Installing Batteries

Details the procedure for inserting and replacing the unit's batteries.

3. Set-up Menu Options

3.2 Setting up Program Test Points and Pass;Fail Limits

Configures test locations and pass/fail thresholds for measurements.

3.4 Establishing ATP Pass;Fail Limits

Provides guidance on setting pass/fail criteria for ATP measurements.

4. Sample Measurements and Test Results

4.1 Taking a Sample Measurement

Step-by-step guide on how to perform a sample measurement using the luminometer.

4.2 Performing a Retest

Instructions for retesting a failed sample measurement.

6. Operator Maintenance

6.3 Cleaning and Replacing the Protective Pocket

7. Troubleshooting

7.2 Troubleshooting Tips

Offers general advice and checks for resolving luminometer issues.

7.3 Luminometer Error Codes

10. Technical Specification

11. Cautions and Warnings

Hygiena SystemSURE Plus v2 Specifications

General IconGeneral
MeasurementRelative Light Units (RLU)
Detection Range0 to 9999 RLU
DisplayLCD
ConnectivityUSB
Battery TypeRechargeable Lithium-ion
Operating Temperature5°C to 40°C (41°F to 104°F)
Instrument TypeATP Monitoring System
TechnologyBioluminescence
Weight0.23 kg (0.5 lb)
Compatible TestsHygiena ATP test devices
Dimensions7.5 x 3.5 x 1.5 inches

Related product manuals