Error messages
33
EN
12 Error messages
12.1 Blood pressure measurement errors
The blood pressure monitor indicates blood pressure measurement errors and communication errors by emitting
12 short acoustic signals.
Error
description
Cause
Remedy
Err 1
Arm moved during
measurement.
Keep your arm still during the measurement (
Chapter 6.3).
Insufficient valid pulse
rate detected.
Put the blood pressure cuff on again (
Chapter 6.2).
Err 2
Arm moved during
measurement.
Keep your arm still during the measurement (
Chapter 6.3).
The blood pressure cuff
is not correctly
positioned on your arm.
Check the placement of the blood pressure cuff (
Chapter 6.2).
Err 3
Blood pressure beyond
the measurement range.
If this notification appears continuously (or repeatedly), it is
possible that the blood pressure monitor is not suitable for you.
Contact your doctor.
Strong arm movement.
Keep your arm still during the measurement (
Chapter 6.3).
Err 4
Readings not
transmitted correctly.
Wait until you take another measurement. The data are then
transmitted together with the new readings.