Biotech-Lab User Manual \\ 30www.ibpmedical.com
HDU-PRS15
Range: - 12 to 15 psi
-0.85 to 1 bar
-650 to +775 mmHg
Resolution: 0.01 mmHg
Accuracy: General 0.1% full scale
Media
Compatibility: Gasses and Fluids
Over pressure: 2 x full scale
HDU-PRS30
Range: -13 to 36 psi
-0.93 to 2.5 bar
-700 to +1900 mmHg
Resolution: 0.01 mmHg
Accuracy: General 0.1% full scale
-500 to 1000 mmHg ± 1
mmHg
Media
Compatibility: Gasses and Fluids
Over pressure: 1.5 x full scale
HDU-PRH50
Range: -13 to 50 psi
-0.93 to 3,5 bar
-700 to 2580 mmHg
Resolution: 0.1 mmHg
Accuracy: 0.05% full scale
Media
Compatibility: Gasses
Over pressure: 2 x full scale
HDU-PRH100
Range: -13 to 100 psi
-0.93 to 7 bar
-700 to 5150 mmHg
Resolution: 0.1 mmHg
Accuracy: 0.05% full scale
Media
Compatibility: Gasses
Over pressure: 2 x full scale
HDU-PRS100
Range: -12 to 100 psi
-0.85 to 7 bar
-650 to 5150 mmHg
Resolution: 0.1 mmHg
Accuracy: 0.1% full scale
Media
Compatibility: Gasses and Fluids
Over pressure: 2 x full scale