Biotech-Lab User Manual \\ 33www.ibpmedical.com
9.4.1 TECHNICAL DATA
Temperature HDU-Pt100 Range: -50...100 °C
Resolution: 0.001 °C
Accuracy: 25 ... 40°C ± 0.015 °C
otherwise ± 0.025 °C
if Sensor and interface
are adjusted together
Calibration PTB/NIST traceable
Communication Rate Five times per second
Degree of Protection IP54
Power Supply 3.3 to 5 V, max. 70 mA
via RS485- or USB-Interface
Environmental Conditions Ambient Operating Temperature Range: 0 to 50°C
Storage Temperature Range: -40 to 80°C
Operating and Storage Humidity: 85% max
relative humidity non-condensing
from 0 to 50°C
Dimensions 64 x 45 x 25 mm
9.5 PT100 SENSORS Temperature Sensor Family: Air - Surface - General Dip - Laboratory
Rugged Pt100 temperature sensor family
Air General - Dip Surface Laboratory
RS485
Interface
Technical data of the sensors are available at:
http://www.ibpmt.com/hdu_e/adoc.asp?hdu_e/hdu_tps.htm.
Pt100
Interface