Biotech-Lab User Manual \\ 37www.ibpmedical.com
12.4 HANDLING OF REFERENCE SOLUTIONS
IBP Medical standard solutions are produced under ISO13485:2003 quality management. They are traceable to NIST and PTB
Standards Reference Materials and are sealed with tamper-evident packaging.
To ensure standard solution and calibration/verification accuracy.
• Keep solutions tightly capped to avoid evaporation
• Do not return used solutions to the storage bottle
• Do not remove solutions from their original bottle
• Keep the solutions in a cool place
• Use only fresh reference solutions for calibration and verification
• Use the solution immediately after pouring, evaporation will cause errors
• Discard solution the appropriate number of days after opening the bottle
• Discard solution after the expiration date