29
WARNING
Check the position of the test lead set in current test
mode. Place the red lead at mA only when below
400mA, and place it at 10A when the value to be tested
is over 400mA, or the value is unknown.
WARNING
Do not attempt to measure voltage when the test leads
are connected to the mA or 10A port. Doing so will
create a short circuit condition and blow the internal fuse
or cause damage to the instrument and possibly cause
bodily injury.