Sample Preparation and Storage
Proper Sample Cup Volume
The volume of plasma, serum, or urine sample required varies based on the number of slides being
used in the run:
Number of slides Sample cup fill volume (µL)
1 60
2 70
3 80
4 90
5 100
6 110
7 120
8 130
9 190
10 200
11 210
12 220
13 230
14 240
15 250
16 260
17 270
18 280
Sample Inspection After Centrifugation
It is good practice to examine the sample carefully following centrifugation in a centrifuge
and/or in the analyzer (by running a whole blood separator). If fibrin strands can be seen in the
sample, they may interfere with sample pipetting. It may be necessary to rim the serum/plasma with
a wooden stick, respin the sample, and proceed.
Various conditions, such as hemolysis, may affect results. You might also want to modify your test
panel based on the following visual observations. Refer to the “Chemistry Descriptions” section on
pages 3051 for information about how each condition may affect specific chemistries.
Note: We recommend that after you have centrifuged a sample in a Catalyst whole blood separator
that you inspect the sample for the conditions listed above.
Hemolysis
Visual: Sample has a transparent reddish hue ranging from pale pink to deep red.
Indications: Damage to red blood cells during sample preparation or intravascular
hemolysis.
Icterus
Visual: Plasma has a transparent yellow to opaque brown color.
Indications: Obstructive or toxic liver disease, intravascular hemolysis.

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Idexx Catalyst One and is the answer not in the manual?

Idexx Catalyst One Specifications

General IconGeneral
TypeVeterinary Chemistry Analyzer
Sample TypeWhole blood, serum, plasma, urine
Humidity10% to 85% non-condensing
ConnectivityUSB, Ethernet
Supported SpeciesCanine, Feline, Equine, Bovine
Weight22 lbs
Operating Temperature15°C to 30°C
Power Supply100-240 VAC, 50/60 Hz
DisplayTouchscreen
Sample Volume100 µL
Result Time8 minutes
Regulatory ComplianceCE

Related product manuals