Quality Control
Overview
The purpose of quality control (QC) is to verify the integrity of your slides and also to verify that your
Catalyst One* analyzer is functioning properly.
You should run a QC test:
When the analyzer is first installed.
After cleaning the internal components of the analyzer.
If the analyzer has been moved.
To verify system performance.
Quality Control Materials
IDEXX VetTrol* Control
In each box of IDEXX VetTrol* Control, there are four vials containing freeze-dried powder (dark
brown bottle marked “VetTrol Control”) and four vials containing diluent (lighter bottles marked
“Diluent for VetTrol”). The lot numbers for the diluent and the control are different and can be found
on the product packaging.
For more information about IDEXX VetTrol Control, see its package insert.
Storage
Control and diluent vials should be stored frozen (-1C/0°F). Discard opened control vials
within 24 hours. Expired or unwanted material should be discarded with other clinical waste.
Note: Do not store in the freezer door; only in the main freezer compartment.
Stability and Handling
For most chemistries, IDEXX VetTrol Control can be used up to 24 hours after reconstitution
when it is stored in the refrigerator and equilibrated to room temperature before running (do not
leave at room temperature for more than 2 hours). For creatine kinase and ammonia values,
IDEXX VetTrol Control fluid should be used within 2 hours following reconstitution. Exposure to
light will affect total bilirubin and creatine kinase results. Ammonia concentration will increase
with time.
UPRO Control
In each box of UPRO Control, there are six vials containing the control fluid. The lot number can be
found on the product packaging.
Storage
Control fluid should be refrigerated (2°C8°C/3F46°F). Discard at the expiration date.
Expired or unwanted material should be discarded with other clinical waste.
Stability and Handling
Use within 24 hours after opening (refrigerate when not in use).

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Idexx Catalyst One and is the answer not in the manual?

Idexx Catalyst One Specifications

General IconGeneral
TypeVeterinary Chemistry Analyzer
Sample TypeWhole blood, serum, plasma, urine
Humidity10% to 85% non-condensing
ConnectivityUSB, Ethernet
Supported SpeciesCanine, Feline, Equine, Bovine
Weight22 lbs
Operating Temperature15°C to 30°C
Power Supply100-240 VAC, 50/60 Hz
DisplayTouchscreen
Sample Volume100 µL
Result Time8 minutes
Regulatory ComplianceCE