26
Group 3 ALB
CREA
Mg
PHOS
BUN/UREA
URIC
UCRE
Group 4 ALT
LDH
Group 5 ALKP
GGT
Group 6 AST
Ca
CK
UPRO (to be used with UPRO Control fluid only)
Preparing Control Fluid
The instructions for preparing control fluid vary depending on the type of control you are preparing.
To Prepare IDEXX VetTrol Control Fluid
1. Remove one diluent and one control vial from freezer. Allow 6090 minutes for vials to
acclimate to room temperature.
2. Slowly invert the diluent vial several times to thoroughly mix the contents. Do not shake.
3. Gently tap the control vial on the counter several times to dislodge any material adhering to
the stopper.
4. Remove the seal and stopper from each vial just before adding the diluent to control. Do not
leave the vials open.
5. Transfer exactly 3.0 mL of diluent to the control vial, using a clean, dry, Class A volumetric
pipette or an equivalent automatic pipette. Discard the remaining diluent.
IMPORTANT: Measurement must be precise or results will be incorrect.
6. Replace the stopper on the control vial and hold it firmly in place. Gently invert the vial 610
times every 10 minutes for one hour (the use of a slow rocker is recommended). Do not
shake. Reconstitution, with occasional inversion, will take 4560 minutes. Visually verify that
all freeze-dried material is dissolved before using.
7. Run quality control on the Catalyst One analyzer (see instructions below).
To Prepare UPRO Control Fluid
1. Take one vial of UPRO Control out of the refrigerator and gently invert it 610 times to mix
thoroughly.
2. Transfer 300 µL of UPRO Control into a Catalyst* sample cup (to be loaded in the sample
drawer).
3. Let the contents in the sample cups reach room temperature (approximately 10 minutes).
4. Run quality control on the analyzer.
Quality Control

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Idexx Catalyst One and is the answer not in the manual?

Idexx Catalyst One Specifications

General IconGeneral
TypeVeterinary Chemistry Analyzer
Sample TypeWhole blood, serum, plasma, urine
Humidity10% to 85% non-condensing
ConnectivityUSB, Ethernet
Supported SpeciesCanine, Feline, Equine, Bovine
Weight22 lbs
Operating Temperature15°C to 30°C
Power Supply100-240 VAC, 50/60 Hz
DisplayTouchscreen
Sample Volume100 µL
Result Time8 minutes
Regulatory ComplianceCE