29
d. Using an IDEXX-supported alcohol prep pad, wipe the
incubator ring and optics window in a counterclockwise
direction. Repeat this step at least three times using a new
alcohol prep pad for each wipe.
e. Clean the white reference tile using a new alcohol prep pad.
f. Using a dry optical tissue, dry the optics window and
reference tile, ensuring all signs of dampness have
evaporated from the cleaned components. If streaks or
smudges remain, repeat the cleaning process.
g. Replace the carousel inside of the analyzer, lower the
carousel cover and close the side door.
h. Tap Done.
Cleaning the Outside of the Analyzer
and the Sample Drawer
Clean the outside of the analyzer or sample drawer with a damp (not wet) lint-free cloth. A mild
liquid soap will remove grease. Do not use any of the following near the analyzer: organic solvents,
ammonia-based cleaning products, ink markers, sprays containing volatile liquids, insecticides,
disinfectant, polish, or room freshener.
Care should be taken not to spill any samples, chemicals, cleaning agents, water, or other fluids
on/in the analyzer.
Note: Dust and animal hair can lead to analyzer failures. Routinely dust off the analyzer with a
damp cloth and dust around its location. Do not block the cooling vents under the analyzer by
allowing paper, loose materials, or dust to accumulate.
WARNING: Never wipe the analyzer or its surroundings with ammonia-based cleaning products.
Avoid urine odors around analyzer. Ammonia in the atmosphere will falsely increase ammonia (NH
3
)
quality control and patient test results.
Emptying the Waste Drawer
It is essential that you empty the waste drawer after every run or when prompted. The analyzer will
not operate when the waste drawer is full. Pull the waste drawer to remove it from the analyzer.
Maintenance
2d
2e

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Idexx Catalyst One and is the answer not in the manual?

Idexx Catalyst One Specifications

General IconGeneral
TypeVeterinary Chemistry Analyzer
Sample TypeWhole blood, serum, plasma, urine
Humidity10% to 85% non-condensing
ConnectivityUSB, Ethernet
Supported SpeciesCanine, Feline, Equine, Bovine
Weight22 lbs
Operating Temperature15°C to 30°C
Power Supply100-240 VAC, 50/60 Hz
DisplayTouchscreen
Sample Volume100 µL
Result Time8 minutes
Regulatory ComplianceCE

Related product manuals