OPERATION MANUAL
IFF-701Ti
1-2-1
Page 2
Jul 1/00
1.4.4 CAUTION and WARNING Labels
Exercise extreme care when performing
operations preceded by a CAUTION or
WARNING label. CAUTION labels appear
where possibility of damage to equipment
exists. WARNING labels denote conditions
where bodily injury or death may result.
1.5 AC POWER REQUIREMENTS
The Test Set operates over a voltage range
of 100 to 120 VAC at 60 Hz or 220 to
240 VAC at 50 Hz according to VOLTAGE
SELECT Switch setting.
The specified fuse ratings are listed in
1-2-1, Table 1.
CAUTION: FOR CONTINUOUS
PROTECTION AGAINST FIRE,
REPLACE ONLY WITH FUSES
OF THE SPECIFIED VOLTAGE
AND CURRENT RATINGS.
INPUT VOLTAGE F1 AND F2 FUSES
100 to 120 VAC 1.0 A, 250 V
Fast Blo (Type F)
(IFR PN: 5106-4501-000)
(1.25” AGC Glass)
220 to 240 VAC 0.5 A, 250 V
Fast Blo (Type F)
(IFR PN: 5106-0000-016)
(1.25” AGC Glass)
Specified Fuse Ratings
Table 1
1.6 FCC REQUIREMENTS
The Test Set is prohibited from use in VAR
(variable) frequency modes during ramp
use. Any test signals radiated from the
antenna must be fixed frequencies. The
Test Set is prohibited from externally
modulating any band during ramp use. This
includes microphones, audio generators,
etc.
1.7 BATTERY RECHARGING
Refer to 1-2-1, Figure 1.
STEP PROCEDURE
1. Verify FUSES are correct for normal
operating voltage. Refer to 1-2-1,
para 1.5.
2. Verify VOLTAGE SELECT Switch is
set to the setting (115 or 230 VAC)
that matches the input ac power.
3. Connect ac power cable between AC
Power Connector and normal
operating voltage power source,
according to Test Set configuration.
Refer to 1-2-1, para 1.5.
NOTE: The battery charger operates
whenever ac power is applied
to the Test Set.
4. Allow several hours for battery charge
or until 0 to 30% scale displays
battery voltage is 15 V or greater.
NOTE: Overnight charging is
recommended.