OPERATION MANUAL
IFF-701Ti
1-2-1
Page 3
Jul 1/00
OUT
NAV TONE
0.5A @ 220VAC
1.0A @ 110VAC
F2
TYPE F
F1
PWR MONSIG GEN
DEMOD DEMOD
115
S
230
00920004
AC IN CONNECTOR
Battery Recharging
Figure 1
1.8 BENCH OPERATION
STEP PROCEDURE
1. Verify FUSES are correct for normal
operating voltage. Refer to
1-2-1, para 1.5.
2. Verify VOLTAGE SELECT Switch is
set to the setting (115 or 230 VAC)
that matches the input ac power.
3. Connect ac power cable between AC
Power Connector and normal
operating voltage power source
according to Test Set configuration.
Refer to 1-2-1, para 1.5.
4. Press PWR/BAT Switch to PWR
position and verify POWER ON
Indicator illuminates.
5. Connect UUT to Test Set.
1.9 EXTERNAL CLEANING
The following procedure contains routine
instructions for cleaning the outside of the
Test Set.
CAUTION: DISCONNECT POWER FROM
TEST SET TO AVOID
POSSIBLE DAMAGE TO
ELECTRONIC CIRCUITS.
STEP PROCEDURE
1. Clean front panel buttons and display
face with soft lint-free cloth. If dirt is
difficult to remove, dampen cloth with
water and a mild liquid detergent.
2. Remove grease, fungus and ground-in
dirt from surfaces with soft lint-free
cloth dampened with isopropyl
alcohol.
3. Remove dust and dirt from connectors
with soft-bristled brush.
4. Cover unused connectors with
suitable dust covers to prevent
tarnishing of connector contacts.
5. Clean cables with soft lint-free cloth.
6. Paint exposed metal surface to avoid
corrosion.