24
Technical data
LR 1000 basic LR 1000 control
Nominal voltage VAC 100 – 120
230
Frequency Hz 50 / 60
Input power W 1200
Viscosity max. mPas 100000
Usable volume ml 300 – 1000
Useful volume with dispersing ml 500 – 1000
Attainable vacuum mbar 25
Stirring speed range rpm 10 – 150
Speed display LED TFT
Speed setting resolution rpm 1
Speed deviation rpm ± 5
Working temperature max. (medium) °C 120
Temperature display LED TFT
Temperature setting resolution K 1 0.1
Temperature measurement resolution K 0.1
Heating control accuracy K ± 1
Connection for external temperature sensor Pt 100
Control accuracy with external sensor K ± 0.2
Adjustable safety circuit °C 47 (± 10) – 225 (± 20)
Type of cooling Liquid through cooling
Minimum temperature of the cooling medium °C 3
Cooling medium permissible pressure bar 1
Weighing range kg - 0 – 2
Weighing resolution g - 1
pH meter interface no yes
pH value display - TFT
pH display range - 0 – 14
pH display resolution - 0.1
Nominal torque Ncm 3
Torque trend display - TFT
Timer function yes
Timer display LED TFT
Interface USB, RS 232
Protection class acc. to EN 60529 IP 21
Permissible ambient temperature °C +5 ... +40
Permissible ambient humidity (relative) % 80
Dimension (W x D x H) mm 443 x 295 x 360
Weight kg 16
Operation at a terrestrial altitude m max. 2000
Subject to technical changes!