22
Error codes
The fault is shown by an error message in the display as following if the error occurs.
Proceed as follows in such cases:
Switch off the device using the main switch
Carry out corrective measures
Restart the instrument
Error code Description Effect Corrective action
Error 3
(Er 3)
Internal temperature of device is too high.
Heating off
Motor off
- Switch off the instrument and allow it to cool down.
Error 4
(Er 4)
Motor is blocked or overloaded. Motor off - Switch off the instrument
- Decrease the load and restart again
Error 10
(Er 10)
Remote control is interrupted. Heating off
Motor off
- Change watchdog time
- Check relating connection
Error 11
(Er 11)
External temperature sensor plugged
during heating control.
Heating off - Unplug the external temperature sensor.
Error 12
(Er 12)
External temperature sensor unplugged
during heating control.
Heating off
- Plug in the external temperature sensor
Error 14
(Er 14)
Short circuit of external temperature
sensor or cable.
Heating off
- Check the external temperature sensor and cable
Error 18
(Er 18)
The temperature of safety sensor is
higher than safety temperature setting.
Heating off
- Increase the safety temperature setting or allow
cooling down
Error 51
(Er 51)
Mains voltage is too high. Heating off
Motor off
- Switch off the device and check the mains voltage.
Error 52
(Er 52)
Mains voltage is too low. Heating off
Motor off
- Switch off the device and check the mains voltage.
Er 60
Power is interrupted when heating or
motor control is activated in Mode B and C
- Exit error state when any button is pressed.
Only for LR 1000 basic!
If the actions described fails to resolve the fault or another error code is displayed then take one of the following steps:
- Contact the service department
- Send the instrument for repair, including a short description of the fault.