10 11
Intended use
///
Use
Use RC 2 lite for tempering and circulating fluids.
Intended use: Tabletop device
Unpacking
›
Please unpack the device carefully.
› Any damage should be notified immediately to the shipping agent (post oce, railway
network or logistics company).
1 RC 2 lite
2 Power cord set
3 USB cable
4 Hexagon socket offset screw key
5 Hose olives NW 8 (2 x)
6 User guide
7 Warranty card
IKA
RC 2
l
ite
Garanty
1+2+3
1
2
6
7
5
4
3
/// Refrigerant
Warning!
› The device is not allowed to be used in an ATEX (Atmosphere Exlosible) zone.
› Flammable refrigerant used. Do not use mechanical devices or other means to accelerate the
defrosting process, other than those recommended by the manufacturer.
› Do not damage the refrigerant circuit.
›
Insufficient ventilation may result in the formation of explosive mixtures. Only use the device
in well ventilated areas and keep all air vents open.
›
Flammable refrigerant used. Only the manufacturer may open and repair the closed refrigeration
system.
› The minimum room size is 9 m³, so that in the event of an error (exit of the complete refrigerant
filling quantity) no flammable refrigerant-air mixture can arise (R 290 max. 8 g/m³ ambient air).
/// Range of use
Indoor environments similar to that a laboratory of research, teaching, trade or industry.
The safety of the user cannot be guaranteed:
›
if the device is operated with accessories that are not supplied or recommended by the
manufacturer.
›
if the device is operated improperly or contrary to the manufacture’s specifications.
›
if the device or the printed circuit board are modified by third parties.