/// USB interface
The Universal Serial Bus (USB) is a serial bus for connecting the device to the PC. Equipped with
USB devices can be connected to a PC during operation (hot plugging).
Connected devices and their properties are automatically recognized.
USB 2.0
type B
or
/// USB device drivers
First, download the latest driver for IKA devices with USB interface from:
www.ika.com/ika/lws/download/usb-driver.zip.
Install the driver by running the setup file. Connect the IKA device through the USB data cable
to the PC. The data communication is via a virtual COM port. Configuration, command syntax
and commands of the virtual COM ports are as described in RS 232 interface.
Note: the USB driver is uncessary for Windows 10 system. Don't install it for Windows 10
system!
Fluids (Standard information for IKA fluid)
No.
IKA Designation
Operating
temperature range
for open bath
application
(°C)
Operating
temperature range
for closed bath
applications
(°C)
Safety
temperature
(°C)
Flash point
(°C)
0
CF.EG28.N10.80.8
-10 ... 80 -10 … 80 90 115
1
CF.EG39.N20.80.16
-20 … 80 -20 … 80 90 115
2
CF.EG44.N25.80.19
-25 … 80 -25 … 80 90 115
3
CF.EG48.N30.80.22
-30 … 80 -30 … 80 90 115
4
UF.Si.N30.150.10LV
-30 … 130 -30 … 150
145
>170
5
HF.Si.20.200.50
20 … 200 20 … 200 255 >280
6
HF.Si.20.250.50A
20 … 200 20 … 250 255 >280
7
Water
5 … 95 5 … 95 - -
8
Customized
(1) Classification:
HF: Heating Fluid
CF: Cooling Fluid
UF: Universal Fluid
(2) Chemical composition:
Si: Silicone oil
EG: Ethylene Glycol
(3) Temperature range: (
Minimum temperature. Maximum temperature)
N: Negative Temperature
(4) Viscosiy:
Viscosiy at 25 °C for Heating Fluid (HF)
Viscosity at -20 °C for Cooling Fluid (CF)
Viscosity at 25 °C for Universal Fluid (UF)
Dynamic viscosity [mPa
s] is a product of kinematic viscosity [mm²/s] and density [kg/m³] of the
fluid divided by 1000.
(5) Additional information:
A: Oil Additives
LV: Low Viscosity
(5) Addition information
(4) Viscosity (8
mm²/s
)
(3) Temperature range (-10 – 80 °C)
(2) Chemical composition (Ethylene Glycol 28%)
(1) Classification (Cooling fluid)
Note: for open bath application!
Note: Tap water may be unsuitable for operation because the calcium carbonate content
may cause calcification.
High purity water (from icon exchangers) and distilled or bi-distilled water are unsuitable for
operation due to corrosive properties of these media. High purity water and distillates are
suitable as a medium after adding 0.1 g soda (Na
2
CO
3
, sodium carbonate) per liter of water.
Note: The temperature limit values are adjustable in accordance with the fluid used.
Check the suitability of the liquid according to your application.
Nomenclature for IKA fluids:
CF.EG28.N10.80.8 --
Interfaces and outputs
The device can be connected to a PC and operated with the laboratory software labworldsoft
®
through USB or RS 232 interface.
The device software can also be updated with a PC via the RS 232 or USB port.
Note: Please comply with the system requirements together with the operating instructions
and help section included with the software.