EasyManua.ls Logo

impact SV005 - Page 14

impact SV005
21 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
13
Sieving intensity
Sieving time
Sieving mode
For a correct set up please follow with great care our instructions:
SIEVING INTENSITY
The Sieving intensity can be varied entering a value from 0 to 99 on the display.
1. Press several times the key until on the numeric display (C8) appears the letter H (set
up mode referred to the sieving intensity).
2. Press the key
. The middle dash of the numeric display (C8) will start lighting.
3. Press the key to increase the value or the key to decrease it. The led is
now blinking showing that the value is under modification.
4. Press the key until the figure on the right is blinking.
5. Press the key to increase the value or the key to decrease it.
6. Press several times the key
until you come back to mode L (residual time to end the
test). Confirming the data storing, the led will stop blinking.
WARNING
The intensity of the sieving must not be set over the values that may cause
the bumping up to the end of scale of the electromagnet (this condition can
be recognized by a marked metallic noise)
If this condition is shown, reduce the intensity of the sieving until the
unusual noise disappears
The non-respect of such shrewdness may cause the braking of some
components of the sieve-shaker (springs and electromagnet) and
successively the decay of the WARRANTY term
WARNING
Be sure to set the intensity of sieving before initiating the screening, or the shaker may not
start.
SIEVING TIME
Setting up a value from 1 to 999 minutes can vary the sieving period on the display.
1. Press several times the key
until on the numeric display (C8) appears the letter t
(indicating the sieving time or the possibility to select it).