and carefully rub the measuring chamber to a bright
finish.
Do not bend the anode.
We recommend to rub the inside walls of the measuring
chamber up to the groove for the retaining ring to a bright
finish using a polishing cloth.
The sealing surfaces must only be worked
concentrically.
Do not bend the anode.
Rinse the measuring chamber with alcohol and allow it to
dry.