tina56e1-f (2015-09) 23
Measurement range 1.57 … 9.05 V
Output signal U [V]
Pressure p
1E+05
1E+04
1E+03
1E+02
1E+01
1E+00
1E–01
1E–02
1E–03
1E–04
1E–05
0.0 0.5 1.0 1.5 2.5 3.5 4.5 5.5 6.5 7.5 8.52.0 3.0 4.0 5.0 6.0 7.0 8.0 9.0
1.57
9.05
9.5
mbar
Pa
Torr
Error
p = 10
(U-c)
⇔
U = c + log
10
p
valid in the range 5×10
-5
mbar <p< 1500 mbar
U p c U p c
[V] [mbar] 5.8751 [V] [micron] 3
[V] [µbar] 2.8751 [V] [Pa] 3.8751
[V] [Torr] 6 [V] [kPa] 6.8751
[V] [mTorr] 3
where p pressure
U output signal
c constant (pressure unit dependent)