21
Step 7Power up the unit and from the Main Menu select Maintenance>Shut Down.
Step 8 At the menu prompt: Remove Foam Can instead, remove the Storage Fluid Line from Can Interface
Port. Cap the adapter fitting and stow the tube set in the notch on the storage bracket as shown in Figure 4.3.
Ensure that the open end of the permeate line is secured in a waste container. Select Continue.
Step 9 Remove the Wash Solution bottle from the Storage Fluid Adapter and cap tightly. (Warning: contents
will spill if the unit is shipped with the Wash Solution bottle installed onto the adapter)
Step 10 At the menu prompt: Place Maintenance Tip into Tip Port. Select Continue (Maintenance Tip is already
in place).
Step 11 – Continue past the menu prompt: Insert Storage Fluid Line into Can Port Select Continue. The canister
port must remain open to allow fluid to be flushed out of the permeate port.
Step 12At the menu prompt: Power Off Unit or Restart. Power Off the instrument.
Step 13 Place the orange Canister Interface Port cap that came with the instrument over the Canister Interface
Port.
Step 14 – Remove and drain the Permeate Line.
Step 15 Use the original Foam Packaging and Box for all transport and shipments.