6.6 RECOMMENDED GENERAL MAINTENANCE FOR CP SELECT
Perform Start Up
Routine Daily
Main Menu > Maintenance > Start Up
Eliminates "salt crystals" in CP Select fluid
lines.
Perform Shut Down
Routine Daily
Main Menu > Maintenance > Shut Down
Eliminates " salt crystals " in CP Select fluid
lines.
Available at www.Innovaprep.com or your
local distributor
Maintains a tight seal to the tips, conditions
the O-rings.
Available at www.Innovaprep.com or your
local distributor
Buildup can occur in the permeate hose after
extended use.
After Approx. 50
Shut Down
and/or Start Up
Routines
Available at www.Innovaprep.com or your
local distributor
Wear over time will cause Maintenance Tips to
fit loosely which may cause leakage or
insufficient uptake of the storage fluid.
Replace Tube Set for
Storage Fluid
Adapter
Available at www.Innovaprep.com or your
local distributor
Plastic components, although durable, may
wear over time. Inspect this tube set regularly
for proper operation during startup and
shutdown routines.
Annual
Maintenance and
Calibration
Contact InnovaPrep or your distributor for
return material authorization.
The pressure transducer should be calibrated
every 12 months.