6.6 RECOMMENDED GENERAL MAINTENANCE FOR CP SELECT
Recommendation
Part Number
Frequency
Tips
Reason
Perform Start Up
Routine Daily
N/A
Before day's use
Main Menu > Maintenance > Start Up
Eliminates "salt crystals" in CP Select fluid
lines.
Perform Shut Down
Routine Daily
N/A
After day's use
Main Menu > Maintenance > Shut Down
Eliminates " salt crystals " in CP Select fluid
lines.
Lubricate O - rings
HC08535
Every 2 weeks
Available at www.Innovaprep.com or your
local distributor
Maintains a tight seal to the tips, conditions
the O-rings.
Replace Permeate
Line
HC08006
As Needed
Available at www.Innovaprep.com or your
local distributor
Buildup can occur in the permeate hose after
extended use.
Replace
Maintenance Tip
HC08005
After Approx. 50
Shut Down
and/or Start Up
Routines
Available at www.Innovaprep.com or your
local distributor
Wear over time will cause Maintenance Tips to
fit loosely which may cause leakage or
insufficient uptake of the storage fluid.
Replace Tube Set for
Storage Fluid
Adapter
HC08613
As Needed
Available at www.Innovaprep.com or your
local distributor
Plastic components, although durable, may
wear over time. Inspect this tube set regularly
for proper operation during startup and
shutdown routines.
Annual
Maintenance and
Calibration
N/A
12 Months
Contact InnovaPrep or your distributor for
return material authorization.
The pressure transducer should be calibrated
every 12 months.