TABLE 2.1 CONCENTRATING PIPETTE SELECTION GUIDE
Part Number
Filter Tip
Media Type
Pore Size
Membrane
Surface Area
cm
2
Input Sample
Volume (Varies
by Matrix)
Final Concentrated
Sample Volume
(User Selectable)
Processing Rate
(Varies by Matrix)
mL/min
Notes
CC08000-10
Flat Membrane
Polycarbonate
Track Etch
0.4 μm
8.5
Up to 1 Liter*
200 - 1000 μL
Up to
100 mL /min.
Recommended for concentrating bacteria,
whole cells, spores, pollen, and parasites from
food and beverage matrices or samples
containing proteins and polysaccharides
CC08001-10
Flat Membrane
Polyethersulfone
0.1 μm
8.5
Up to 1 Liter*
200 - 1000 μL
Up to
100 mL /min.
Recommended for concentrating bacteria,
whole cells, spores, pollen, and parasites from
environmental samples
CC08018-10
Hollow Fiber
Polysulfone
0.45 μm
98
Up to 5 Liters*
150 - 1000 μL
Up to
150 mL/min
Recommended for concentrating bacteria,
whole cells, spores, pollen, and parasites from
high fouling matrices
CC08022-10
Hollow Fiber
Polysulfone
0.2 μm
98
Up to 5 Liters*
150 - 1000 μL
Up to
150 mL/min
Recommended for concentrating bacteria,
whole cells, spores, pollen, and parasites from
all sample types
CC08020-10
Hollow Fiber
Polysulfone
0.05 μm
98
Up to 3 Liters*
150 - 1000 μL
Up to
90 mL/min
Recommended for virus concentration. The
addition of Tween 20 to the sample is
recommended to aid recovery efficiency (add
10% solution of Tween 20 in a 1/100 ratio to
the sample)
CC08003-10
Hollow Fiber
Polysulfone
Ultra -
Filtration
98
Up to 500 mL*
150 - 1000 μL
50 mL/min with
Tween 20
* With regards to starting sample volume, it is recommended that the entire sample should be processed within 30 minutes. It has been observed that there is a reduction
in the recovery when processing times exceed 30 minutes. The recommended upper-end run volume as shown in the concentrating pipette tip selection guide and time of
30 minutes is a general recommendation and may be dependent on sample matrix and target organism. See Section 10 for full part number list.