10.0 Technical Information—Specifications
Size 8.2cm(H) x 6.1cm(W) x 2.1cm(T)
Weight 87.9 g
Power Source One 1.5 V Alkaline AA battery
Maximum Current 200 mA at 500 ohm; 121 mA at 1 K ohm
Maximum Voltage 121 V at 1 K ohm; <150 V at 1 M ohm
Number of Operation Modes 2 - Exercise, Walking
Number of Channels 1
Pulse Type Asymmetrical Biphasic
Pulse Width 25-300 microseconds (Adjustable)
Maximum Stimulation Period 3 seconds
Stimulation Trigger Source Tilt or Heel Sensor
Controls and Indicators ON/OFF/Intensity; Stimulation,
Exercise
Error
Shipping and Device (Long Term)
Storage Conditions: Temperature: -4° – 140°F (-20° – +60° C)
Humidity: 95% max., non-condensing
Electrodes (Long Term)
Temperature: 41° – 80.6°F (5° – 27° C)
Humidity: 35 – 50%
Electrodes (Short Term - less than 1 month)
Temperature: 32° – 104° F (0° – 40° C)
Humidity: 35 – 50%
Operating Conditions: 0° – 40° C
Altitude less than 2400 meters (8000 feet)
IEC 60529 Rating: IP21
Expected Service Life: 5years
Do not hesitate to contact your clinician should you have any
questions regarding this device and its safe operation and use.
If further assistance is required, please contact:
Innovative Neurotronics, Inc.
Austin, TX
888-884-6462 www.walkaide.com