Chapter 2: Specifications
22 M47-17041-EN
Stroke 305 mm (12 in) 305 mm (12 in)
(C) Column Diameter 76 mm (3 in) 89 mm (3.5 in)
(D) Maximum Overall Height Refer to the appropriate General Assembly drawing provided in the “Reference
Manual-Equipment”
5
Overall Frame Area (width (E) x depth (F))
J1 743 x 994 mm (29 x 39.125 in) 920 x 1207 mm (36.25 x 47.5 in)
J3 1039 x 994 mm (40.875 x 39.125 in) 1250 x 1207 mm (49.25 x 47.5 in)
Weight Refer to the appropriate General Assembly drawing provided in the “Reference
Manual-Equipment”
5
Speed Information
Testing Speed Range
(at full load)
0.1 to 150 mm/min (0.004 to 6 in/min) 0.1 to 76 mm/min (0.004 to 3 in/min)
Jog (Adjusting) Speed
(Maximum)
150 mm/min (6 in/min) 76 mm/min (3 in/min)
Position Information
Accuracy
6
C4
Equal to or less than +/-0.13 mm (0.005 in) or +/-0.1% of displayed reading,
whichever is greater
Resolution
7
C4 1 µm (0.00004 in)
Cyclic Test and Frequency Limits Capable of tension/tension or compression/compression cyclic testing only. Cyclic
tests are limited to ramp type waveforms as defined within the materials testing
software package being used with the system. Frequency limits are defined by the
compliance of the specimen and the dynamic control limit of the system. For cyclic
tests based on a load rate, the load rate should not exceed a rate equivalent to the
capacity of the frame in a 30 second period. Some variation in this loading rate will
be experienced based on specimen compliance. Although this may be used as a
guideline, further limitations will be encountered if the loading rate creates a speed
requirement near the maximum speed of the hydraulic cylinder. In all cases, the
frequency for cyclic testing is limited to 1 Hz or less.
Center Alignment
8
(estimated) Estimated axial alignment between the actuator rod and central thread in the load
cell with crosshead in lowest position is 0.64 mm (0.025 in)
Power Requirements 200-250 VAC, 3Phase, 50/60 Hz, 15 Amp
380-415 VAC, 3Phase, 50/60 Hz, 10 Amp
440-480 VAC, 3Phase, 50/60 Hz, 10 Amp
Certifications Conform to all relevant European standards and carry a CE mark
Table 5. LX frame technical specifications. (Continued)
Specification
1
300LX 600LX