Model DX Static Hydraulic
Universal Testing System
Operating Instructions
M47-17027-EN Revision C
The difference is measurable
®

Table of Contents

Other manuals for Instron DX Series

Questions and Answers

  • D
    Daniel MartinezAug 16, 2025
    How to troubleshoot Instron DX frame not receiving power?
    • A
      Allison RobertsonAug 16, 2025
      If your Instron Test Equipment frame isn't receiving power, start by checking the incoming power supply to ensure it's properly connected and functioning. If the power supply is working, a fuse might be blown. Inspect all system fuses and, if necessary, consult the manual for instructions on replacing them.
  • M
    Maria GreerAug 19, 2025
    What to do if HPS will not start on Instron DX Test Equipment?
    • R
      robertdavisSep 12, 2025
      If the HPS on your Instron Test Equipment won't start, make sure the Emergency Stop button is disengaged. Enable the frame and then try starting the HPS again. Also, check that the system disconnect switch is set to ON ( | ). Another potential cause could be a tripped limit switch. Inspect the limits, adjust the frame if needed to reset the limit, and then enable the frame before starting the HPS.
  • E
    emily95Aug 22, 2025
    How to fix communication lost between frame and controls on Instron DX?
    • A
      Alex RamosAug 22, 2025
      If you've lost communication between the frame and controls on your Instron Test Equipment, try resetting the system. Follow the shutdown and startup procedures outlined in the System Operating Instructions.
  • O
    ochapmanAug 25, 2025
    Why is position measurement not working on Instron DX?
    • E
      erneststewartAug 26, 2025
      If the position measurement isn't working on your Instron Test Equipment, and the piston is at the bottom of the cylinder, enable the frame, start the HPS, and move the piston up approximately 6 mm (0.25 in). Then, zero all measurements and calibrate all transducers. Also, check the position encoder cable for any damage or loose connections.
  • S
    stephensteresaAug 27, 2025
    What to do when communication is lost with the user control panel on Instron Test Equipment?
    • S
      Susan RobinsonAug 27, 2025
      If you've lost communication with the user control panel on your Instron Test Equipment, check the user control panel cable for any signs of damage.
  • J
    James OrtizAug 30, 2025
    How to enable the frame on Instron Test Equipment?
    • A
      Amy HudsonAug 30, 2025
      If the frame is not enabled on your Instron Test Equipment, enable it through the software interface. For Partner systems, go to Machine and then Enable Frame. For Bluehill systems, select the Frame icon and then Enable Frame. If you are using Partner software, check the “Enable Frame Timer” setting in Tools/Configure/Controllers; ensure the “Enable Frame Timer” box is NOT checked.
  • T
    Tonya MurilloSep 2, 2025
    Why is the specimen slipping in the grip jaws of my Instron DX?
    • S
      Scott WhiteSep 2, 2025
      If the specimen is slipping in the grip jaws of your Instron Test Equipment, unload the specimen and reclamp the grip jaws. Ensure the correct grip jaw type is being used for the specimen. Inspect the grip jaws for damage. For frames with G1-style crossheads, use additional force to clamp the grip jaws (but do not hammer grip components). For frames with G7-style crossheads, ensure the pressure gauge reads approximately 20.7 bar (300 psi) while clamping.
  • G
    gonzalesamandaSep 5, 2025
    What to do if the grip jaws are improperly lubricated on Instron DX Test Equipment?
    • J
      Joel StewartSep 5, 2025
      If the grip jaws are improperly lubricated on your Instron Test Equipment, unload the specimen and remove it from the grips. Relubricate the grips, assemble the grip components, and then install the specimen in the grip jaws.
  • D
    Deborah MercadoSep 7, 2025
    How to troubleshoot strain instrument not working on Instron Test Equipment?
    • J
      John OliverSep 7, 2025
      If the strain instrument on your Instron Test Equipment may not be working properly, check the instrument for proper operation and ensure that the zero pin was removed from the instrument.
  • D
    Denise ZamoraSep 10, 2025
    How to reset adjustable crosshead limit on Instron DX?
    • J
      Justin CarterSep 10, 2025
      If the adjustable crosshead limit has been tripped on your Instron Test Equipment, use the adjustable crosshead control switch to move the adjustable crosshead in the opposite direction. This will reset the limit switch.

Instron DX Series Specifications

General IconGeneral
SoftwareBluehill Universal
Frame TypeDual column
ApplicationsTensile, compression, flexure
Frame StiffnessHigh stiffness for accurate measurements
Test Speed Range1000 mm/min
Displacement Resolution0.0001 mm
Load Measurement Accuracy+/- 0.5% of reading
Control Modesdisplacement, strain, position
Data Acquisition RateUp to 2 kHz
Power Requirements220-240 V, 50/60 Hz
WeightVaries by model
Load Capacity5 kN to 500 kN (1, 125 lbf to 112, 500 lbf) depending on model