Chapter 3: Operation
34 M47-17027-EN
8. Remove the 4 set screws and lock nuts (No. 2 in Figure 6).
9. Reinstall the rear and side panels. Reconnect the power supply and power on the machine. Check for
leaks and proper machine operation.
Procedure for 600DX frames
1. Power off the machine and disconnect the power cord.
2. Remove the rear and side panels from the base.
3. Using a 6 mm hex wrench, remove the mounting hardware (six black socket head cap screws from
the left and right of the machine (No. 1 in Figure 7).
4. Using a 10 mm hex wrench, lower the leveling feet on the base of the HPS (No. 2 in Figure 7) so that
the HPS is supported on the leveling feet.
5. Check clearances all around the base of the HPS. If any part of it is in contact with the frame base,
use the leveling feet to make adjustments to avoid contact.
6. Reinstall the rear and side panels. Reconnect the power supply and power on the machine. Check for
leaks and proper machine operation.
Mount fixtures
Warning
Figure 7. Location of screws on 600DX
Always wear appropriate personal protective equipment when preparing, operating and
maintaining this equipment. Personal protective equipment should include, but is not
limited to, eye protection and head protection. Other types of personal protective
equipment may also be needed. You must perform your own risk assessment and take
appropriate measures to protect yourself from harm.