Service manual
EyeSeeCam vHIT
4
2. Maintenance and replacing parts
2.1 Maintenance
If the surface of the goggle is contaminated, it is cleaned with a soft cloth moistened with a mild solution of water
and dish washing liquid or similar. The use of organic solvents and aromatic oils must be avoided.
Always disconnect the USB cable during the cleaning process and keep any fluid from entering the inside of the
instrument or the accessories.
Before cleaning always switch off and disconnect from power.
Use a soft cloth lightly dampened with cleaning solution to clean all exposed surfaces.
Do not allow liquid to enter the camera housing.
Do not autoclave, sterilize or immerse the instrument or accessory in any fluid.
Do not use hard or pointed objects to clean any part of the instrument or accessory.
RECOMMENDED
CLEANING AND
DISINFECTION SOLUTION
Warm water with mild, nonabrasive cleaning solution (soap).
SERVICE &
MAINTENANCE
The vHIT system must be serviced at least once a year.
PROCEDURE
Clean the instrument by wiping the outer case with a lint free cloth
lightly dampened in cleaning solution.
SPECIFIC PARTS
Goggle: the hot mirrors and the silicone tubes should be cleaned
regularly.
Lens: should be cleaned on demand. For this, use appropriate tissue
to avoid scratches or damage.
Cleaning the silicone tubes
Make sure the device is disconnected from the
power supply before starting this procedure.